Will an ACQUITY Sample Manager start an injection sequence with the door open? - WKB20867
Article number: 20867
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Fixed Loop (FL)
- ACQUITY H-Class Sample Manager - Flow Through Needle (FTN)
- ACQUITY I-Class Sample Manager - FTN
- ACQUITY I-Class Sample Manager - FL
- ACQUITY Systems with Waters Driver Pack 4 or greater
ANSWER
The ACQUITY Sample Manager (Autosampler) will NOT initiate an injection sequence if it detects the door is open.
This means that it will pause and wait if the door is open and it has not started the injection. Once it has started, such as bringing the needle down into a well, the system is committed to completing the cycle and it will bring the needle to the seal and complete the injection.
Note: The instrument will move slowly and trigger a beep to warn that this movement is in progress.
ADDITIONAL INFORMATION
