ACQUITY BSM, SM, CM system overpressure error - WKB208750

Article number: 208750

SYMPTOMS

  • One or more column manager positions cause system overpressure error
  • Without column in place, pressure is as expected
  • With PEEK tubing in place of column system, overpressure occurs

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC BSM
  • ACQUITY UPLC Column Manager

CAUSE

Blocked column outlet tubing

FIX or WORKAROUND

Replace the column outlet tubing (700004161).