SYMPTOMS
- One or more column manager positions cause system overpressure error
- Without column in place, pressure is as expected
- With PEEK tubing in place of column system, overpressure occurs
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC Column Manager
CAUSE
Blocked column outlet tubing
FIX or WORKAROUND
Replace the column outlet tubing (700004161).