ACQUITY UPLC Sample Manager fails needle seal leak test after PM - WKB208621
Article number: 208621
SYMPTOMS
- ACQUITY UPLC Sample Manager fails needle seal leak test after PM
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Faulty injection valve
FIX or WORKAROUND
Replace the injection valve (700002765).
ADDITIONAL INFORMATION
id208621, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR