Overpressure on H-Class system - WKB207419

Article number: 207419

SYMPTOMS

  • Overpressure    

ENVIRONMENT

  • H-Class

CAUSE

Tubing between FTN and Column Manager is blocked.

FIX or WORKAROUND

Replace or try to flush the tubing.

ADDITIONAL INFORMATION

(Part number for tubing between FTN and column manager = 700005476)