2489 Dual Absorbance Detector - times out to idle mode - WKB207374
Article number: 207374
SYMPTOMS
- Shutter closes while user observes real-time response, even though instrument is not under software control
- "Idle" mode seen
ENVIRONMENT
- 2489 Dual Absorbance Detector
CAUSE
The shutter closes by default.
FIX or WORKAROUND
In Configuration, clear the selection of "disable shutter idle mode".

ADDITIONAL INFORMATION