Article number: 204075

SYMPTOMS

  • "Assertion Failed...[random characters]" error appears at the beginning of detector initialization sequence

ENVIRONMENT

  • 2489 UV/Vis Detector
  • Firmware is current, v1.80

CAUSE

The gray ribbon cable connecting the Front End Analog PCB to the Personality PCB (J25) is loose or disconnected.

FIX or WORKAROUND

Reseat the cable connections.