"Wavelength verification failure" error occurs during initial diagnosis with ACQUITY UPLC PDA Detector - WKB204090

Article number: 204090

SYMPTOMS

  • “Wavelength verification failure” error during the initial diagnosis

ENVIRONMENT

  • ACQUITY UPLC PDA Detector

CAUSE

Air in the cell or UV absorption from solvent in cell

FIX or WORKAROUND

  1. Turn on the PDA detector with the flow on in advance.
  2. If the solvent currently filling the PDA detector cell is a solvent miscible with methanol,
    turn on the detector with methanol running.

ADDITIONAL INFORMATION

id204090, A-30PDA, UPAPDA, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA