Pressure fluctuates on ACQUITY UPLC BSM Binary Solvent Manager - WKB20289

Article number: 20289

SYMPTOMS

  • Pressure is stable for a period of time but then starts bouncing

ENVIRONMENT

  • ACQUITY
  • ACQUITY UPLC BSM (Binary Solvent Manager)

CAUSE

Air in the system

FIX or WORKAROUND

  1. Prime each line individually.
  2. Flush the system with 100% methanol while maintaining a back pressure of at least 10,000 psi for at least one hour before returning to the required mobile phases.

ADDITIONAL INFORMATION