Liquid seen on Alliance vials - WKB202852
Article number: 202852
SYMPTOMS
- Liquid on Alliance vials
- Compression check failure
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Injector was not purged.
FIX or WORKAROUND
Purge the injector.