What is the system volume of the HPLC and UPLC systems? - WKB201368

Article number: 201368

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • Arc HPLC System
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Arc

ANSWER

The system volumes of the Waters Alliance HPLC, Arc HPLC and the ACQUITY UPLC series are as follows.

 

・Alliance e2695 Separations Module : <650 µL (independent of backpressure @ 1.0 mL/min.)

・Acquity UPLC (BSM+FL) : <120 μL (Including a 50 μL mixer)

・Acquity H-Class (QSM+FTN) : <400 uL (Including a 100 μL mixer)

・Acquity I-Class (BSM+FTN) : <100 uL (Including a 50 μL mixer)

・Acquity I-Class (BXM+FL) : <95 μL (Including a 50 μL mixer)

ACQUITY Arc (QSM-R+FTN-R) : Path 1 ≦ 1450 μL, Path 2 ≦ 1150 μL (Gradient delay volume : Path 1 ≦ 1050 μL, Path 2 ≦ 750 μL)

・Arc HPLC (QSM-R+FTN-R) : ≦ 1350 μL (Gradient delay volume : ≦ 1000 μL)

 

ADDITIONAL INFORMATION

・Waters e2695 Separations Modules

 https://www.waters.com/webassets/cms...20002552en.pdf

・ Next Generations Instrument Design(Acquity UPLC)

 https://www.waters.com/webassets/cms...cs/wa42889.pdf

・ACQUITY UPLC in the Chromatography Lab: Analytical Considerations

 https://www.waters.com/webassets/cms...cs/wa42886.pdf

・ACQUITY UPLC H-Class System

 https://www.waters.com/webassets/cms...20003294en.pdf

・Waters UPLC I-Class System - Pinnacle of UPLC Performance

 https://www.waters.com/webassets/cms...UPLC_Intro.pdf

・ACQUITY Arc System

 https://www.waters.com/webassets/cms...20005400en.pdf

・Arc HPLC System Instrument Specifications

 https://www.waters.com/webassets/cms...20006861en.pdf

 

 

 

 

Not able to find a solution? Click here to request help.