Article number: 201298

SYMPTOMS

  • Organizer console error stops batch analysis
  • Condensate is collected but does not flow to the waste bottle

ENVIRONMENT

  • ACQUITY Sample Organizer

CAUSE

The condensate drain tube in front of the Sample Organizer is not draining condensate into the waste container.

FIX or WORKAROUND

Reconfigure the Organizer drain tube so that the condensate drains from the Organizer into the solvent waste container via gravity flow.

ADDITIONAL INFORMATION