ACQUITY UPLC Sample Organizer gives "Organizer Transfer Shuttle homing error" - WKB201352

Article number: 201352

SYMPTOMS

  • Plate is not removed from shelf completely
  • Error: "Organizer Transfer Shuttle homing error"
  • Mostly lower self positions in the sample organizer (position 2 and 3) are affected. The plates and shelves for these positions don't sem to line up correctly when transferring
  • All other shelves operate normally

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer
  • ACQUITY UPLC Common Platform Sample Organizer (CPSO)

CAUSE

  • Damaged Igus E-chain for ACQUITY UPLC Sample Organizer
  • Damaged Igus E-chain for CPSO

FIX or WORKAROUND

Replace the Igus E-chain.

  • Sample Organizer Igus E-chain part number 700002984
  • CPSO E-chain part number 700005751

ADDITIONAL INFORMATION