FTN Rotary Tray not calibrated and Z-axis calibration out of range - WKB200830
Article number: 200830
SYMPTOMS
Multiple errors generated
- FTN Rotary Tray not calibrated
- After the rotary tray calibration, the same error appears when the analysis starts
- “Z-axis calibration out of range” appears with “Characterize Needle Seal”.
- The area around the needle does not go down with the replacement needle.
- Zp-axis move h/w fault
- Communication Failure
ENVIRONMENT
- FTN Sample Manager
CAUSE
Needle Carriage damaged / Ribbon cables damaged
FIX or WORKAROUND
- Replace the Needle carriage (700005225)