Error: Failed to calibrate system volume - WKB200904
Article number: 200904
SYMPTOMS
- Cannot characterize needle and loop volumes - Error occurs around 2.8 minutes.
- All console leak tests pass
- Air sensor calibration passes
- New needle and loop have been correctly installed, volumes are correct and system is well primed.
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Puncture pin is blocked - air bubble cannot be drawn through tubing to measure system volumes
FIX or WORKAROUND
- Close Masslynx.
- Remove plates.
- Power off sample manager and manually move the needle carriage to front left of the sample manager.
- Remove needle from carriage and leave hanging in sample compartment.
- Use T6 tool to remove the puncture pin - loosen grub screw and slide the puncture pin downwards, out of carriage assembly.
- Clean puncture pin physically with thin tool, or sonicate in solvent.
- Reassemble needle carriage.
- Power on Sample manager.
- Open MassLynx.
- Run Characterize needle and loop volumes in console.
ADDITIONAL INFORMATION
id200904, UPBINARY