2707 leaks during syringe priming - WKB198908
Article number: 198908
SYMPTOMS
- 2707, without sample heater/cooler, leaks from condensed water and leakage tubing (right hand-side waste tubing) during syringe priming
ENVIRONMENT
- 2707 Sample Manager
CAUSE
The waste tube from the syringe drive is not inserted into the drip tray. This can occur after replacing the syringe drive (700003839).
FIX or WORKAROUND
Remove the lower-front cover and insert the syringe drive waste tubing into the drip tray.
ADDITIONAL INFORMATION
id198908, 2707, 2707SFC