2707 leaks during syringe priming - WKB198908

Article number: 198908

SYMPTOMS

  • 2707, without sample heater/cooler, leaks from condensed water and leakage tubing (right hand-side waste tubing) during syringe priming        

ENVIRONMENT

  • 2707 Sample Manager

CAUSE

The waste tube from the syringe drive is not inserted into the drip tray. This can occur after replacing the syringe drive (700003839).

FIX or WORKAROUND

Remove the lower-front cover and insert the syringe drive waste tubing into the drip tray.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.