UPLC peak splitting - WKB198916
Article number: 198916
SYMPTOMS
- UPLC peak is extremely broad and splits in two
ENVIRONMENT
- ACQUITY UPLC System
- Partial loop mode
CAUSE
Wrong solvent bottle for the weak wash line.
FIX or WORKAROUND
Use the mobile phase of initial conditions as the weak wash.