Stepping seen in the chromatography with 2489 UV/Vis detector - WKB198137

Article number: 198137

SYMPTOMS

  • Stepping seen in the baseline

ENVIRONMENT

  • 2695 Separations Module
  • 2489 UV/Vis Detector

CAUSE

Leaking flow cell

FIX or WORKAROUND

 Rebuild or replace the flow cell. 

ADDITIONAL INFORMATION