Primary pump passes dynamic leak test with negative leak rate values - WKB196173
Article number: 196173
SYMPTOMS
- Successful dynamic leak test
- Primary pump head passes dynamic leak test with negative leak rate values
- Primary pump head occasionally passes with acceptable leak rate values
- Increase of leak rate on accumulator seen when leak rate on primary is monitored
- Seals and check valves are new, replaced more than once on both primary and accumulator pumps
ENVIRONMENT
- ACQUITY UPC2 ccBSM
- ACQUITY UPLC I-Class BSM
- ACQUITY Classic
CAUSE
Faulty primary pump head
FIX or WORKAROUND
- Replace the primary pump head.
- Perform the dynamic leak test.
ADDITIONAL INFORMATION
High pressure seal marks might be seen on primary pump head.