Alliance 2695 failing static leak test after passing with pin plug installed at transducer - WKB19616
Article number: 19616
SYMPTOMS
- Failing static leak test.
- "Accumulator check valve is okay but accumulator leak rate is too high..." error is displayed.
- Passing Static Leak Test with pin plug installed at accumulator transducer.
- Failing Static Leak Test with pin plug reinstalled at the in-line filter.
- Failing Static Leak Test with pin plug reinstalled at accumulator transducer.
ENVIRONMENT
- 2695 Separations Module
- PM performed recently
CAUSE
Accumulator transducer is faulty.
FIX or WORKAROUND
1. Replace accumulator transducer, Pressure Transducer [WAT270966]
2. Perform static leak test.
ADDITIONAL INFORMATION
id19616, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, checkvalve