Alliance 2695 failing static leak test after passing with pin plug installed at transducer - WKB19616

Article number: 19616

SYMPTOMS

  • Failing static leak test.
  • "Accumulator check valve is okay but accumulator leak rate is too high..." error is displayed.
  • Passing Static Leak Test with pin plug installed at accumulator transducer.
  • Failing Static Leak Test with pin plug reinstalled at the in-line filter.
  • Failing Static Leak Test with pin plug reinstalled at accumulator transducer.

ENVIRONMENT

  • 2695 Separations Module
  • PM performed recently

CAUSE

Accumulator transducer is faulty.

FIX or WORKAROUND

1. Replace accumulator transducer, Pressure Transducer [WAT270966]

2. Perform static leak test.

ADDITIONAL INFORMATION