How to disassemble an ACQUITY UPLC QSM - WKB193250
Article number: 193250
OBJECTIVE or GOAL
Shut down and disassemble a QSM for shipment to the repair center.
ENVIRONMENT
- ACQUITY UPLC QSM
PROCEDURE
- Wash the system with H2O for 20 minutes, and then wash the system with IPA for 20 minutes.
- Remove the solvent filters from the solvent bottle and place them in a plastic bag (as shown below).
- Close the control software and power-off all modules on the stack.
- Electrically unplug all UPLC modules and mark and remove any cables (Ethernet, signal, I/O, events) from the modules.
(Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of the cable and tubing are also helpful.)
- Remove fluidics connections:
- Green PEEK tube, Purge solvent
- Syringe waste line, gray tube
- Needle wash waste line (corrugated tube)
- Inline filter/mixer
- Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
Note the dimensions and weight of the module to choose the right box:
ACQUITY UPLC QSM 34.3 cm X 66.1 cm X 23.8 cm, 27.5 kg - Fill and sign the latest revision of the H&S module provided by Waters.
-
Attach it to the box and send a PDF copy to the local Waters office.
-
The Waters office will contact you to arrange the shipment.
ADDITIONAL INFORMATION
To reinstall the module, attach all of the cables and tubes that were removed.
id193250, A-30SDM, isopropanol, UPAQSM, UPPQSM, UPQSM, UPQSM+, UPQSMBIO