SYMPTOMS

  • Retention time varies

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

The seal wash plunger seal is leaking inside the pump head.

FIX or WORKAROUND

Replace the plunger seal (WAT271018).

Find details on the replacement of these parts in the Alliance e2695 Separations Module Operators Guide.

ADDITIONAL INFORMATION

As seal damage may be due to worn/ scored plungers (WAT270959) It may be prudent to replace them at the same time as the seals.