QSM fails dynamic leak test - WKB192601

Article number: 192601

SYMPTOMS

  • Leak test fails for accumulator pump head

ENVIRONMENT

  • ACQUITY UPLC QSM

CAUSE

Faulty tubing between the accumulator pump head transducer outlet and vent valve inlet port 3

FIX or WORKAROUND

Replace the tubing (430002316) between the accumulator pump head and the vent valve.

ADDITIONAL INFORMATION