Sample manager makes grinding noise when it starts - WKB192589
Article number: 192589
SYMPTOMS
- Sample manager makes grinding noise when it starts
- X-axis H/W error in MassLynx
ENVIRONMENT
- ACQUITY UPLC H-Class Bio
CAUSE
The seat port assembly tubing is in the way of the needle carriage.
FIX or WORKAROUND
Push the seat port assembly tubing away from the needle carriage travel path.
ADDITIONAL INFORMATION
id192589, A-30SM, audible noise, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+