2998 gives baseline ripple at lower UV wavelengths - WKB191137

Article number: 191137

SYMPTOMS

  • Baseline ripple at low UV wavelengths
  • At UV wavelengths higher than 215 nm, the baseline ripple disappears
  • Without mobile phase flowing by the flow cell, the baseline ripple disappears

ENVIRONMENT

  • 2998

CAUSE

An incorrect stroke volume is configured in the instrument method.

FIX or WORKAROUND

Set the stroke volume to an appropriate value, as shown in the chart below.

clipboard_e7b10fbd1f549d9cfc24f870c279a7ee7.png

ADDITIONAL INFORMATION

PDA Baseline Noise (Engineer Troubleshooting) - wkb95692

High delta psi and pressure unstable with e2695 Separations Module - WKB25194

id191137, 2487, 2489, 2489B, 2489SFC, 2996, 2998, 2998B, 2998SFC, A-10PDA, A-10UV, eluent, pressure fluctuation, pressure ripple