Skip to main content
Waters

Interpump communications HW failure on the ACQUITY I-Class BSM - WKB17129

Article number: 17129

SYMPTOMS

  • Interpump communications HW error in console
  • The error is intermittent; occurs every few hours 

ENVIRONMENT

  • ACQUITY UPLC I-Class BSM
  • ACQUITY UPLC M-Class

CAUSE

Corrupt ICS and firmware

FIX or WORKAROUND

  1. Remove the ICS for the BSM and reload the ICS.
  2. Reflash the firmware on the BSM.

ADDITIONAL INFORMATION

Pump ran overnight with flow from both A and B. Error did not return.

id17129, comms, UPBINARY, UPBSM+, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM

Not able to find a solution? Click here to request help.