Skip to main content
Waters

Chromatogram baseline is high on UV detector - WKB17173

Article number: 17173

SYMPTOMS

  • Chromatogram baseline is in higher than normal

ENVIRONMENT

  • 2489 UV/Vis Detector

CAUSE

Poor solvents are leaving dirt in the system.

FIX or WORKAROUND

  1. Remove the column and replace it with a union.
  2. Put A and B lines from the ACQUITY into the alcohol or magic mix, which is 25/25/25/25 water/isopropyl alcohol/methanol/acetonitrile and 0.2% formic acid (can go up to 1% formic acid).
  3. Flush the system by priming for five minutes, and then pump for at least two hours (preferably overnight) at a flow rate that gives at least 6,000-psi pressure.
  4. Prime and flush with fresh LC/MS-grade or Milli-Q water. 
  5. Put on fresh LCMS-grade solvents that are used for the LCMS analysis. If the solvent bottles have not been cleaned recently, use the alcohol mix to clean out the bottles, and then rinse them well with fresh solvent. 
  6. Attach a new column.
  7. Prime and flush out the water using the fresh solvents.
  8. Run the LCMS analysis again. 

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.