Skip to main content
Waters

ACQUITY Sample Manager needle bends after a few injections - WKB1685

Article number: 1685

SYMPTOMS

  • Bent needle after a few injections

ENVIRONMENT

  • ACQUITY Sample Manager
  • UNIFI 

CAUSE

The wrong plate was defined when specifying the sample tray in the analysis.

FIX or WORKAROUND

Ensure that the correct sample tray is configured in the analysis.
Steps to determine this are outlined in UNIFI online Help and article wkb120231.

ADDITIONAL INFORMATION

 

id1685, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR

Not able to find a solution? Click here to request help.