ACQUITY Sample Manager needle bends after a few injections - WKB1685
Article number: 1685
SYMPTOMS
- Bent needle after a few injections
ENVIRONMENT
- ACQUITY Sample Manager
- UNIFI
CAUSE
The wrong plate was defined when specifying the sample tray in the analysis.
FIX or WORKAROUND
Ensure that the correct sample tray is configured in the analysis.
Steps to determine this are outlined in UNIFI online Help and article wkb120231.
ADDITIONAL INFORMATION
id1685, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR