2487 Detector not communicating in Empower - WKB16250
Article number: 16250
SYMPTOMS
- 2487 says "NO" for Status OK in node properties
ENVIRONMENT
- Empower 3 enterprise
- LAC/E
- 2487 Dual Wavelength Detector
- 2695 Separations Module
- Breeze
CAUSE
Incorrect IEEE address.
FIX or WORKAROUND
- Set the IEEE address to the correct address.
- Select Shift, and then DIAG to access the configuration page.
- Navigate to the IEEE address field and specify the desired address.
- Return to the Node properties and select Scan.
ADDITIONAL INFORMATION