Alliance 2695 system fails RSD specification - WKB16262
Article number: 16262
SYMPTOMS
- RSD is failing Height in SQT
- Injection volumes are accurate
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Bubbles in system or inadequate purging injector
FIX or WORKAROUND
- Prime and purge solvent lines and injector several cycles.
- Perform compression check, verify it passed.
- Rerun standards to retest the %RSD.
ADDITIONAL INFORMATION
id16262, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB