Does the ELSD detector need a BPR (back pressure regulator)? - WKB15358

Article number: 15358

ENVIRONMENT

  • ACQUITY ELSD
  • 2424 Evaporative Light Scattering Detector
  • Back pressure regulator (BPR)

ANSWER

No, there is no liquid outflow from the ELSD, so there is no place to connect a back pressure regulator. 

ADDITIONAL INFORMATION

The back pressure regulators (BPRs) should only be used on the last detector in the flow path (exception is the ELSD).