Ethernet instruments "No" in the "OK" column of Node Properties - WKB1476
Article number: 1476
SYMPTOMS
- Connection to system lost
ENVIRONMENT
- Empower Personal
CAUSE
Network Bridge reverted to the defaulted IP address 169.254.xxx.xxx.
FIX or WORKAROUND
- Check the DHCP Window for the IP address of the Ethernet instruments.
- Ping the IP address of the card in the command prompt window.
- Run ipconfig/all in the command prompt verify IP address.
- Delete Bridge.
- Recreate Bridge with the correct IP address for the card.
- Reboot the PC.
- Reboot the instruments.
- Ping card and instrument IP addresses.
- Run as normal.