Article number: 1476

SYMPTOMS

  • Connection to system lost

ENVIRONMENT

  • Empower Personal

CAUSE

Network Bridge reverted to the defaulted IP address 169.254.xxx.xxx.

FIX or WORKAROUND

  1. Check the DHCP Window for the IP address of the Ethernet instruments.
  2. Ping the IP address of the card in the command prompt window.
  3. Run ipconfig/all in the command prompt verify IP address.
  4. Delete Bridge.
  5. Recreate Bridge with the correct IP address for the card.
  6. Reboot the PC.
  7. Reboot the instruments.
  8. Ping card and instrument IP addresses.
  9. Run as normal.

ADDITIONAL INFORMATION

 