With 515/1515/1525 pump: low pressure, leak, or noisy baseline - WKB14760
Article number: 14760
SYMPTOMS
- Low system pressure
- Pump is leaking solvent
- Solvent is dripping from chassis
- No visible source of leak
- Noisy baseline
ENVIRONMENT
- 515 HPLC Pump
- 1515 Isocratic HPLC Pump
- 1525 Binary HPLC Pump
CAUSE
Leaking high-pressure filter assembly (pulse dampener) behind front panel
FIX or WORKAROUND
- Replace the high-pressure filter assembly. 700002742 (515, High Pressure Filter Assy).