Seal pack leaking in a 717 Autosampler - WKB14458

Article number: 14458

SYMPTOMS

  • Seal pack leaking after replacing

ENVIRONMENT

  • 717+ Autosampler

CAUSE

  • The ferrule is set in a position which is flushed with the top of the needle which created space and leak.
  • Seals are not calibrated 

FIX or WORKAROUND

  1. Adjust the ferrule 
  2. Perform seal pack calibration. See the attached 717 operator's manual.