ACQUITY UPLC M-Class does not go into trapping mode - WKB14223
Article number: 14223
SYMPTOMS
- System goes straight from inject into the analytical cycle, even though the method clearly has trapping selected
- Trap valve does not change position
- Trap flow rate is not applied
- Method never says trapping
ENVIRONMENT
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- Xcalibur
- MassLynx
CAUSE
-Corrupt LC method
- Bug of the Mclass driver
FIX or WORKAROUND
- Go into the instrument method.
- Select File "New" and create an entirely new LC/MS method.
- Verify that the trapping method runs properly.
- If is a bug issue, power cycle the MClass for 20 minutes and restart the PC.
- If step 4 doesn't work, recreate DHCP.
ADDITIONAL INFORMATION
Do not use Save As because it merely propagates the error.