Skip to main content
Waters

2424 ELSD shows high baseline noise - WKB14109

Article number: 14109

SYMPTOMS

  • High baseline noise

ENVIRONMENT

  • 2424 ELSD
  • MassLynx

CAUSE

Restricted exhaust

FIX or WORKAROUND

  1. Disconnect the exhaust from existing extraction.
  2. Reconnect to a suitable extraction system.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.