2424 ELSD shows high baseline noise - WKB14109
Article number: 14109
SYMPTOMS
- High baseline noise
ENVIRONMENT
- 2424 ELSD
- MassLynx
CAUSE
Restricted exhaust
FIX or WORKAROUND
- Disconnect the exhaust from existing extraction.
- Reconnect to a suitable extraction system.
ADDITIONAL INFORMATION
id14109, 2424, 2424B, baseline ripple, MLYNX, MLYNXV41, SUPMM