2695 Separations Module "lost prime" - WKB124978

Article number: 124978

SYMPTOMS

  • "Lost prime"

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • e2695 Separations Module

CAUSE

The black release ring is out of position. The plunger is loose in the plunger drive assembly.

FIX or WORKAROUND

  1. Remove the pump head and manifold.
  2. Put the black release ring in the proper position.

ADDITIONAL INFORMATION

 