Column Manager - Active power supply trips uninterruptible power supply (UPS) - WKB12495

Article number: 12495

SYMPTOMS

  • The main circuit breaker trips when ACQUITY System is powered-on
  • After resetting circuit breaker, breaker trips again as soon as the UPS is connected to main power
  • Unplugging the load (ACQUITY System) from UPS enables normal power-up
  • Process of elimination shows that the column manager module trips the UPS

ENVIRONMENT

  • ACQUITY Column Manager - Active (CM-A)
  • ACQUITY Auxiliary Column Manager - Active (CM-Aux)

CAUSE

Faulty power supply

FIX or WORKAROUND

Replace the CM-A power supply (700005433 - POWER SUPPLY, CM-A).