Column Manager - Active power supply trips uninterruptible power supply (UPS) - WKB12495
Article number: 12495
SYMPTOMS
- The main circuit breaker trips when ACQUITY System is powered-on
- After resetting circuit breaker, breaker trips again as soon as the UPS is connected to main power
- Unplugging the load (ACQUITY System) from UPS enables normal power-up
- Process of elimination shows that the column manager module trips the UPS
ENVIRONMENT
- ACQUITY Column Manager - Active (CM-A)
- ACQUITY Auxiliary Column Manager - Active (CM-Aux)
CAUSE
Faulty power supply
FIX or WORKAROUND
Replace the CM-A power supply (700005433 - POWER SUPPLY, CM-A).