2695 Separations Module: The pump flow rate is normal but the retention time varies - WKB124495

Article number: 124495

SYMPTOMS

  • When run with 1 mL/min, the peak usually appears at 6 min, but it gradually elutes faster from 4.5 min to 4 min.
  • Draw the liquid with a syringe in front of the column to check the flow rate → No problem
  • Tried two columns of the same type, the same trend was seen
  • Same trend seen after repreparing the mobile phase
  • The pressure(1300psi) is the same as before, Δpsi is stable with ~13.
  • When the column, sample, and mobile phase are transferred and measured with another same system, the peak elutes at 6 min without any problem.

ENVIRONMENT

  • 2695 Separations Module
  • 2489 UV/Vis Detector
  • A line is used for premix, isocratic, and mobile phase
  • 20% methanol used for B, C, D line

CAUSE

There is a possibility that there is a GPV open on a line other than A, and liquids from other lines are mixed from there.

FIX or WORKAROUND

  1. Prime all lines except A in mobile phase before run, if retention time becomes normal, replace the GPV.

ADDITIONAL INFORMATION

id124495, 2489, 2489B, 2489SFC, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, A-10UV, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent, gradient proportioning valve