2695 Separations Module: The pump flow rate is normal but the retention time varies - WKB124495

Article number: 124495

SYMPTOMS

  • When run with 1 mL/min, the peak usually appears at 6 min, but it gradually elutes faster from 4.5 min to 4 min.
  • Draw the liquid with a syringe in front of the column to check the flow rate → No problem
  • Tried two columns of the same type, the same trend was seen
  • Same trend seen after repreparing the mobile phase
  • The pressure(1300psi) is the same as before, Δpsi is stable with ~13.
  • When the column, sample, and mobile phase are transferred and measured with another same system, the peak elutes at 6 min without any problem.

ENVIRONMENT

  • 2695 Separations Module
  • 2489 UV/Vis Detector
  • A line is used for premix, isocratic, and mobile phase
  • 20% methanol used for B, C, D line

CAUSE

There is a possibility that there is a GPV open on a line other than A, and liquids from other lines are mixed from there.

FIX or WORKAROUND

  1. Prime all lines except A in mobile phase before run, if retention time becomes normal, replace the GPV.

ADDITIONAL INFORMATION

id124495, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent, gradient proportioning valve