H-Class CM-A column outlet tubing leaks - WKB123927
Article number: 123927
SYMPTOMS
- There is a leak in the tubing from the column outlet to the outlet valve of the CM-A
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC Column Manager - Active
CAUSE
Bending of the tubing when installing a column. There may be damage inside the tubing.
FIX or WORKAROUND
Replace the tubing (700005478 - Welded Tube Assy, SS, 19.0 LG).
ADDITIONAL INFORMATION
For ACQUITY UPLC H-Class Bio: 700005483 - Welded Tube Assy, MP35N, 19.0 LG, LP