ACQUITY Pump Error: Pump HW over pressure - WKB123647

Article number: 123647

SYMPTOMS

  • Approximately 2.5k psi back pressure when outlet of mixer disconnected (50/50 H20/MeOH, 1 mL/min)

ENVIRONMENT

  • ACQUITY Binary Solvent Manager (BSM)

CAUSE

Partially blocked mixer

FIX or WORKAROUND

Replace the mixer (700002911 - Classic BSM, 50 µL). 

ADDITIONAL INFORMATION