Skip to main content
Waters

How to reset an Alliance 2695 CPU PCB - WKB12347

Article number: 12347

OBJECTIVE or GOAL

Fully reset the Alliance CPU board.

ENVIRONMENT

  • 2695 Alliance HPLC system

PROCEDURE

  1. Power-on the instrument.
  2. Wait until you hear three beeps or see the Alliance logo on the LCD.
  3. Press and release Clear, and then press and release 0 on the keypad.
  4. The display states that it is going to factory defaults and to power-down the instrument.

ADDITIONAL INFORMATION

Clear 0 function clears all methods, the seal pack calibration, the configuration, and all logs, including the Trace log.

 

Not able to find a solution? Click here to request help.