ACQUITY UPLC Failing Dynamic Leak/Static Decay Test - WKB12345

Article number: 12345

SYMPTOMS

  • Retention time shift
  • High delta psi
  • Baseline ripple
  • Failing dynamic leak test (DLT)
  • Poor reproducibility
  • Replicate injections steadily increasing in volume
  • The static leak test fails, with various messages
  • Excessive leakage
  • Unable to compress or maintain pressure
  • Leak rate above acceptable limit
  • Check valve okay, but excessive leakage

ENVIRONMENT

  1. ACQUITY UPLC BSM
  2. ACQUITY UPLC I-Class BSM
  3. ACQUITY UPLC H-Class QSM
  4. ACQUITY UPLC H-Class BSM
  5. ACQUITY UPLC H-Class Bio BSM
  6. ACQUITY UPLC H-Class Bio QSM
  7. ACQUITY UPLC M-Class BSM
  8. nanoACQUITY UPLC BSM
  9. ACQUITY Premier BSM
  10. ACQUITY Premier QSM
  11. Arc Premier QSM
  12. ACQUITY APC

CAUSE

Failed seals in the head(s)

FIX or WORKAROUND

Replace plunger and seals:

  • Plunger: 700002600  

  • Seal Wash Housing Seal: H-Class/ACQUITY UPLC - 700002598, M-Class - 700009838, H-Class Bio/ACQUITY Premier - 700005422, I-Class / APC - 700006048
  • High Pressure Seal:

    • 700002599 ACQUITY UPLC BSM, ACQUITY H-Class QSM

    • 700005418 ACQUITY H-Class Bio, ACQUITY Premier BSM and ACQUITY Premier QSM

    • 700009135 ACQUITY I-Class BSM / Acquity APC

    • 700009838 ACQUITY UPLC M-Class        

Also consider replacing the pump head (700002595 for H-Class, 700005411 for H-Class Bio/ACQUITY Premier and 700009190 for I-Class/APC) and the seal wash housing (700002597 for H-Class, 700005410 for H-Class Bio/ACQUITY Premier and 700009194 for I-Class) 

Note: If seal wash housing shows a "cross" or is colored it should be replaced

  • Re-run DLT and verify it is now passing
  • Set flow rate and make sure pressure and delta are stable and within acceptable limit                                                                           

ADDITIONAL INFORMATION

Can also use pump head rebuild kits consisting of plungers, seals and check valves.

For ACQUITY UPLC and ACQUITY UPLC H-Class:

  • ACQUITY UPLC Accumulator Pumphead Rebuild Kit: 700004746

i2V Valve Rebuild Kit: 700004747 (for BSMs older than serial number F07UPB593M use Passive Check Valve Rebuild Kit 700004748)

  • If seals, plungers, seals, check valves, heads & seal wash housings have been replaced and failure continues, replace the head to transducer SS tubing (430001120)

For ACQUITY UPLC I-Class:

  • ACQUITY UPLC I-Class Accumulator Pumphead Rebuild Kit: Part Number 700010584
  • ACQUITY UPLC I-Class i2V Primary Pumphead Rebuild Kit: 700010583
  • ACQUITY UPLC I-Class BSM Pump head: 700009190
  • ACQUITY UPLC I-Class BSM Seal wash housing: 700009194
  • ACQUITY UPLC I-Class BSM Head to Transducer tubing Part Number 430002472

Specific for ACQUITY UPLC H-Class Bio/ Acquity Premier:

  • ACQUITY UPLC Titanium / H-Class Bio Seal wash housing: 700005410
  • ACQUITY UPLC I-Class / Premier Accumulator Pumphead Rebuild Kit: 700010582
  • ACQUITY UPLC I-Class / Premier i2V Primary Pumphead Rebuild Kit: 700010581

Example of etched seal housing 

seal housing.jpg

Not able to find a solution? Click here to request help.