ACQUITY UPC2 BSM fails dynamic leak test - WKB123256
Article number: 123256
SYMPTOMS
- Primary DLT failure (Pumps A and B)
- Accumulator DLT failure (Pumps A and B)
ENVIRONMENT
- ACQUITY UPC2 cBSM
CAUSE
Poor pump head seals and/or check valves
FIX or WORKAROUND
Replace the UPC2 pump seals and/or check valves.
ADDITIONAL INFORMATION
In severe cases, a UPC2 pump head will be required.
Replace the appropriate check valve. (Primary 700002596 or Accumulator 700002968 / 700005164)
If the dynamic leak test still fails, replace seals and plunger in the appropriate head (2x WAT076152, 1x 700009136,
1x 700009137, 1x 700009000, will rebuild 2 heads)