ACQUITY UPC2 BSM fails dynamic leak test - WKB123256
Article number: 123256
SYMPTOMS
- Primary DLT failure (Pumps A and B)
- Accumulator DLT failure (Pumps A and B)
ENVIRONMENT
- ACQUITY UPC2 cBSM
CAUSE
Poor pump head seals and/or check valves
FIX or WORKAROUND
Replace the UPC2 pump seals and/or check valves.
ADDITIONAL INFORMATION
In severe cases, a UPC2 pump head will be required.
id123256, checkvalve, UPBINARY, UPBSM+, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM+, UPPBSM