Pressure pulses with 515 pump - WKB123301

Article number: 123301

SYMPTOMS

  • Pressure pulses with 515 pump

ENVIRONMENT

  • 515 pump

CAUSE

One head is not holding pressure.

FIX or WORKAROUND

  1. Use the leak test to determine which head is failing to hold pressure.
  2. Tighten the check valve housings and fittings.
  3. If the problem remains, replace the check valves, seals, and plunger.
  • WAT022934 - Plunger Seal Replacement Kit (Clear-100, 1/pk) 
  • WAS207069 - Plunger Assembly 
  • 700000254 - PerformancePLUS CV Cartridges, R/S, 1/8-in

ADDITIONAL INFORMATION