Poor %RSD for ACQUITY UPLC, though it passes sample syringe leak test - WKB123188

Article number: 123188

SYMPTOMS

  • Poor peak area %RSD for ACQUITY UPLC (0.5%-1.0%)
  • Passes sample syringe leak test
  • Passes characterize needle and loop volume three times, with 0 μL difference

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC Sample Manager 

CAUSE

Syringe developed an uneven spot, below where it stops to run the leak test.

FIX or WORKAROUND

Replace the sample syringe.