Skip to main content
Waters

Poor %RSD for ACQUITY UPLC, though it passes sample syringe leak test - WKB123188

Article number: 123188

SYMPTOMS

  • Poor peak area %RSD for ACQUITY UPLC (0.5%-1.0%)
  • Passes sample syringe leak test
  • Passes characterize needle and loop volume three times, with 0 μL difference

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC Sample Manager 

CAUSE

Syringe developed an uneven spot, below where it stops to run the leak test.

FIX or WORKAROUND

Replace the sample syringe.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.