Skip to main content
Waters

Low sensitivity with PDA - WKB123038

Article number: 123038

SYMPTOMS

  • Low sensitivity on LC-UV system
  • Read energy - 25,000 at 230 with water in flow cell
  • Injections look like blanks (500 µL in 2-mL vial)
  • Peak is present when injecting 2 mL in 2-mL vial

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager - FTN
  • ACQUITY UPLC PDA Detector

CAUSE

Sample Manager requires z-axis calibration.

FIX or WORKAROUND

Calibrate Z-Axis per WKB28509.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.