Low sensitivity with PDA - WKB123038
Article number: 123038
SYMPTOMS
- Low sensitivity on LC-UV system
- Read energy - 25,000 at 230 with water in flow cell
- Injections look like blanks (500 µL in 2-mL vial)
- Peak is present when injecting 2 mL in 2-mL vial
ENVIRONMENT
- ACQUITY UPLC H-Class Sample Manager - FTN
- ACQUITY UPLC PDA Detector
CAUSE
Sample Manager requires z-axis calibration.
FIX or WORKAROUND
Calibrate Z-Axis per WKB28509.