What is the part number of a lamp for a 2489, ACQUITY PDA, ACQUITY TUV, or 2998? - WKB1191

Article number: 1191

ENVIRONMENT

  • 2489 UV/Vis Detector 
  • 2998 Photodiode Array Detector
  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC Tunable UV Detector
  • ACQUITY UPC2 PDA Detector
  • ACQUITY Premier PDA eLambda Detector
  • ACQUITY Premier TUV 
  • Thermally Managed ACQUITY PDA or TUV (All)

ANSWER

Note: 700009330 is the lamp within PM kit 201000281 and 700005269 is the lamp within PM kit 201000273. Clickhere for more information

ADDITIONAL INFORMATION

The replacement part number for 201000186 is 201000281 - ACQUITY PDA/TUV 2489/2998 PERFORMANCE MAINTENANCE KIT .

  • If the serial number of your TUV is K05UPT699N or earlier, then the PM kit (lamp) part number is WAS081142
  • If the serial number of your TUV has a high brightness lamp and is later than K05UPT699N, then the PM kit (lamp) part number is 201000281 (does not have intelligent chip)
  • If the serial number of your TUV is A10TUV001N or later, then the PM kit (lamp) part number is 201000273 (has intelligent chip for thermally managed systems)