What causes an ACQUITY UPLC H-Class APH to become clogged? - WKB119150
Article number: 119150
ENVIRONMENT
- ACQUITY UPLC H-Class
ANSWER
It may be the mobile phase, sample preparation, the mobile phase replacement between runs, or the end method of the run.
The precautions are as follows:
- Filter both mobile phase and sample with a 0.2-µm filter.
- Do not flow the organic solvent directly on the line where the buffer passed; wash with water in between.
- If system will not be used for a while, wash off the buffer with water and replace it with organic solvent.
ADDITIONAL INFORMATION
id119150, Active Preheater, eluent, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO