What causes an ACQUITY UPLC H-Class APH to become clogged? - WKB119150

Article number: 119150

ENVIRONMENT

  • ACQUITY UPLC H-Class

ANSWER

It may be the mobile phase, sample preparation, the mobile phase replacement between runs, or the end method of the run.
The precautions are as follows:

  • Filter both mobile phase and sample with a 0.2-µm filter.
  • Do not flow the organic solvent directly on the line where the buffer passed; wash with water in between.
  • If system will not be used for a while, wash off the buffer with water and replace it with organic solvent.

ADDITIONAL INFORMATION

 

id119150, Active Preheater, eluent, UPBINARY, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO