ACQUITY PDa produces curved baseline - WKB118907

Article number: 118907

SYMPTOMS

  • Customer reports a known-method curved baseline
  • Read energy low (below 10000 counts)

ENVIRONMENT

  • ACQUITY PDa
  • ACQUITY UPLC
  • ACQUITY I-Class
  • ACQUITY H-Class

CAUSE

Dirty or damaged flow cell        

FIX or WORKAROUND

  1. Exchange the flow cell.
  2. Increase read energy (minimum > 20000 counts).
  3. Perform test injections.

ADDITIONAL INFORMATION

N/A