ACQUITY I-Class/H-Class/Arc SM-FTN sample draw rate is excessive - WKB118923
Article number: 118923
SYMPTOMS
- Sample drawing rate is excessive
- Multiple errors in error log
- Poor area RSD
ENVIRONMENT
- ACQUITY UPLC I-Class
- ACQUITY H-Class
- ACQUITY Arc
CAUSE
Partially blocked needle
FIX or WORKAROUND
Replace the needle.
ADDITIONAL INFORMATION